1 CG-15005-01 C 3800 North Mill Road · Vineland, NJ 08360 · USA Ph: 1-800-843-1794 · Fx: 1-800-922-4361 Web: www.chemglass.com 120 VAC Power Contro
2 Table of Contents: page Features ………………………………………………………………………….3 General Use ……………………………………………………………………..3 Setup & Operation ……………………
3 Features: 1. A simple turn of a knob adjusts power. 2. Lighted ON/OFF Switch. 3. Circuit protection. 4. 10-Amp output. 1200 Watts Maximum 5. Lo
4 Setup & Operation: 1. Connect the heater / heating mantle to the load receptacle (located on the rear of the unit). 2. Plug the power cor
5 Specifications: Input Voltage: 120VAC, 50/60 Hz. Output Range: 5% to 100% of rated voltage. Maximum Output Power: 1200 Watts +/-5%. Maximum Outpu
6 Maintenance: 1. Do NOT connect heating mantles or heaters (resistive loads) that have power ratings of greater than 1200 watts. 2. Do NOT conne